Abstract
The objective of this study was to determine the result reproducibility and performance of the BD Onclarity human papillomavirus (HPV) assay (Onclarity) on the BD Viper LT platform using both contrived and clinical specimens. Reproducibility was assessed in BD SurePath liquid-based cytology (LBC) medium (SurePath) using contrived panels (HPV genotype 16 [HPV16] positive, HPV18 positive, or HPV45 positive) or clinical specimens (HPV16, -18, -31, -33/58, -45, or -52 positive or HPV negative). In addition, specimens from 3,879 individuals from the Onclarity trial were aliquoted prior to or following cytology processing and tested for HPV. Finally, specimens were collected using either the Cervex-Brush or Cytobrush (or Cytobrush/spatula) for comparison of HPV results. Contrived specimens showed >95% concordance with the expected results, and pooled clinical specimens had standard deviations and coefficients of variation ranging from 0.87 to 1.86 and 2.9% to 5.6%, respectively. For precytology and postcytology aliquot analyses, specimens showed >98.0% overall agreement and mean differences in cycle threshold (CT ) scores for HPV ranging from -0.07 to 0.31. Positivity rates were close between the Cervex-Brush and Cytobrush/spatula for all age groups tested. Onclarity results are reproducible and reliable, regardless of sample collection before or after cytology aliquoting. Onclarity performs well regardless of the method of specimen collection (Cervex-Brush or Cytobrush/spatula) for cervical cancer screening.
Recommended Citation
Young S, Vaughan L, Yanson K, Eckert K, Li A, Harris J, Ermel A, Williams JA, Al-Ghoul M, Cammarata CL, Taylor SN, Luff R, Cooper CK, Van Der Pol B. Analytical and Clinical Sample Performance Characteristics of the Onclarity Assay for the Detection of Human Papillomavirus. J Clin Microbiol. 2020 Dec 17;59(1):e02048-20. doi: 10.1128/JCM.02048-20. PMID: 33087435; PMCID: PMC7771472.